Table of Contents

Advertisement

Technical Description
Technical Data
Table and patient
tabletop
Table tilt:
Table height:
Tabletop:
Patient weight:
Longitudinal travel:
Transverse travel:
Tube assembly stand:
SID:
Oblique projection:
Tube assembly swivel:
6
/ 12

Unit

Motor driven, + 90° to - 17°; Tilting speed approx. 4.5°/s;
min. tilting time 90° to 0°: 21 s (incl. soft start and braking);
automatic stop in horizontal position (0°) can be switched on/off
In horizontal position 89 cm with I.I. 33 or I.I. 40
Dimensions: outside 210 cm x 80 cm, radiotransparent 193 cm x 55 cm
attenuation equivalent:
≤ 0.6 mm Al at 100kV 2.7mm Al HVL
≤ 0.65 mm Al at 100kV 3.7mm Al HVL
Maximum of 200 kg with the following restrictions:
– 181 kg to 200 kg: patient lying centrally on the horizontal, not projecting
tabletop; no table longitudinal and transverse movements permitted;
– 151 kg to 180 kg: patient lying centrally on the tabletop, restriction of the
movement speeds max. 20% and of the table longitudinal movement +/-40
cm and transverse movement to the front 7.5 cm; Trendelenburg position
max. 45° with patient secured with shoulder supports and foot restraints.
– 0 kg to 150 kg: no restrictions of the table longitudinal and transverse
movement. In addition 50 kg for cardiac massage (CPR) in the table mid-
position.
The permitted nominal patient weights were defined with 4 times the test load
according to IEC 601.
160 cm motor-driven, 80 cm each to the head and foot end; speed 6 cm/s (50 Hz)
or 7.2 cm/s (60 Hz)
35 cm motor-driven, 17.5 cm each to the left and right; speed 5 cm/s
115 cm, 150 cm, can be set by motor drive; adjustment at approx. 5 cm/s
Max. ± 40° (SID 115 cm) or ± 35° (SID 150 cm), at max. 8°/s;
fulcrum height adjustable by motor drive 10 to 300 mm above tabletop (= parallax
correction of the object);
automatic parallax compensation between cassette and image intensifier input
screen in the central ray (only with cassette spotfilm device)
Manually in the range of + 90° to - 90° with stops every 10° and -90° to -180° with
stops every 30°
AXD3-340.620.13.01.02