Tomographic sequence

◆ Press the exposure release button up to the first pressure point.
– The tube assembly stand moves to the head-end tomography starting posi-
tion (at the left with the unit horizontal).
◆ Give the breathing command to the patient.
◆ Press the exposure release down fully and keep it pressed until the tube as-
sembly stand stops in the foot-end position.
– During the exposure the tube assembly stand moves towards the foot end
and automatically remains there following the exposure.
With the exposure release button pressed, the tube assembly stand moves back
into the starting position.
Prepare the next tomographic exposure:
◆ Set a new tomographic height.
◆ Press the exposure release button up to the first pressure point.
– The tube assembly stand moves from the foot-end position towards the
head-end tomography starting position (at the left with the unit horizontal).
AXD3-340.620.13.01.02