TV frame rate
Brightness control
TV standard
Dynamic response
TV matrix
TV frame rate
Brightness control
TV standard
Monitor
Tube assembly
Focal spot nominal value
(ICE 60336):
Nominal power (IEC
60613):
Cold power:
Nominal voltage (IEC
60613):
Anode angle:
Anode heat storage ca-
pacity:
Anode rotation frequen-
cy:
Mean continuous pow-
er:
Total filtration (IEC
60601-1-3):
AXIOM Iconos R200
max. 25 f/s; with CAREVISION 12,5; 8; 3 pulses/s
Constant image brightness through automatic gain control
72 Hz progressive scan, independent of frequency
VIDEOMED DH
High-resolution TV camera with maintenance-free 1K CCD sensor for digital
fluoroscopy, fluororadiography and DSA
max. 69 dB, optimized for DSA
Matrix 1024²
max. 30 f/s; with CAREVISION 15; 7,5; 3 pulses/s
Constant image brightness through automatic gain control
72 Hz progressive scanning, independent of frequency
SIMOMED HM; high-line with flicker-free display 72 Hz; 44 cm or.. 54 cm screen
diagonal;
SIMOMED HC; 1024 lines, flicker-free display 72 Hz; 44 cm screen diagonal;
LCD monitor 18"; 46 cm screen diagonal;
OPTITOP 150/40/80HC-100, cover with fan.
0.6/1.0
40 kW/80 kW
52 kW/103 kW
150 kV
12°
580 000 J (783 000 HU)
Radiography: 150 Hz/ 180 Hz
Fluoroscopy: 20 to 30 Hz
max. 450 W
2.5 mm Al at 80 kV
AXD3-340.620.13.01.02
Technical Description
Technical Data