Remove The Covers From Compression Unit - Siemens MAMMOMAT 300 Parts Replacement

Compression mechanism and potentiometer r863
Table of Contents

Advertisement

Procedure - Compression mechanism

Remove the covers from compression unit

System OFF.
Pry open the lid on the maximum compression force presetting knob. Loosen the screw
underneath the lid and remove the knob (1/Fig. 1).
Fig. 1
Remove the cover from the manual compression/decompression knobs (2/Fig. 1).
Release the locking pin on the upper and lower cover and remove the covers.
Loosen the Allen screws on top and at the bottom of the compression unit holding the left
side cover (3/Fig. 1). Remove the side cover.
Siemens AG
Medical Solutions
Locking pin
3
3
SPB7-230.841.03
Rev. 02
04.04
2
1
2
Page 1 of 6
M300/1000/3000/3000 Nova
CS PS 24
2 - 1
2