Scanning Process; End Of The Scene; Organ Programs; Creation Of Programs - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

AXIOM Iconos R200

Scanning process

The X-ray system stands in the starting position for the scanning process.
◆ Press an exposure release button and keep it pressed up to the end of the ex-
amination, i.e. until the scanning process is ended.
◆ Trigger the injection, automatically or manually
◆ Deflect the joystick up / down or press the key to the left / right until the entire
examination area has been scanned.
◆ If necessary reduce the frame rate in a maximum of 2 steps, as filed in the
organ program (saving dose and storage capacity).
Follow the contrast medium flow on the monitor.
Adapt the movement speed of the X-ray system to the contrast medium flow.

End of the scene

◆ Let the exposure release button go.
The exposure series is also ended if the image memory is full and the maximum
scene time has been reached, independently of the position in which the system
is located.

Organ programs

All parameters necessary for the exposure are stored in organ programs.
➩ Refer to the FLUOROSPOT Compact Operator Manual and the POLYDOROS
SX register for further details on the organ programs and their programming.

Creation of programs

➩ Refer to the FLUOROSPOT Compact Operator Manual.

Temporary changes of programs

For the following parameters stored in the organ program, it is possible to change
them temporarily with the corresponding operating elements:
– Film density
– Dominants
– Cu prefilter
– Automatic level 1, 2, 3
The changed parameters apply only until you select another / same organ pro-
gram or switch the system off.
AXD3-340.620.13.01.02
Examination
Digital Radiography
33
/ 52