Dose Reduction - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

Examination
Fluoroscopy
12
/ 52

Dose reduction

In fluoroscopy with CAREVISION (pulsed fluoroscopy) or SUPERVISION the fol-
lowing dose reductions configured in the factory result related to standard fluor-
oscopy:
Automatik-
CAREVISION
programm
15 P/s
1
- 50%
*
2
- 33%
3
- 30%
*
Deviating factory setting for USA
1
for image receptor input dose rate
AXD3-340.620.13.01.02
1
CAREVISION
CAREVISION
7,5 P/s
3 P/s
- 75%
- 90%
- 55%*
- 75%*
- 50%
- 65%
SUPER-
VISION
- 50%
- 50%
- 50%
Operator Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents