Shoulder Supports - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

Application
Attachment
Removal
AXIOM Iconos R200

Shoulder supports

1
To support the patient during examinations in Trendelenburg position,
e.g. myelography
At a Trendelenburg position >15°, the shoulder supports must be used . The max-
imum weight of the patient must not exceed 180 kg to 45° Trendelenburg posi-
tion.
In Trendelenburg position >45°, the max. patient weight of 150 kg must not be
exceeded.
◆ Move the patient table to the horizontal position;
◆ Position the patient;
◆ Loosen the cap screws (1) on both supports;
◆ Move the supports into position on both accessory rails;
◆ Pull the shoulder pads out up to the middle of the patient's shoulders;
◆ Tighten the cap screws (1);
◆ Check the shoulder supports;
◆ Loosen the cap screws (2) for the shoulder pads;
◆ Adjust the height of the shoulder pads for secure and comfortable positioning;
◆ Tighten both cap screws (2).
Warning
When in Trendelenburg position, the patient is protected against slipping only
by the shoulder supports. Pull on the shoulder supports to be sure that they are
securely attached before moving the patient into the Trendelenburg position.
◆ Move the patient table into the horizontal position;
◆ Loosen the cap screws (1) on both sides;
◆ Pull the shoulder supports out of the accessory rails.
AXD3-340.620.13.01.02