Select 1-Point Technique - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

POLYDOROS SX 65/80
Integrated Control Console
Selection
Deselection
Exposure data
to be set
Display of exposure
data
22
/ 36

Select 1-point technique

❏ Setting the kV value
❏ Exposure cut-off by automatic exposure control.
Iconos DFR
Iconos
Iconos
Cassette
Cassette
45
kV
0.3
mA
0.0
0.1
min
0
%hu
H
3.4
µGy*cm 2
◆ Actuate
– a button for film-screen combination or
– a measuring field (dominant) button
With organ program:
◆ Select an organ program with 1-point technique
◆ Actuate
– the mAs button: 2-point technique is selected or
– the ms button: 3-point technique is selected or
– the IONTOMAT button: 3-point technique with IONTOMAT is selected
With organ program:
◆ Select an organ program without 1-point technique
❏ The following exposure data can be set:
– kV value
– Focal spot
– Tube load
– Measuring fields (dominants)
– Film-screen combination
– Film density compensation
❏ After the end of the exposure, the mAs and ms values in the display are flash-
ing.
AXD3-340.620.13.01.02
Iconos Bed
Vertix
ms
40
kV
U
D
80%
80%
100%
Info
mAs
ms
Program
Operator Manual
0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents