Examination
Digital Radiography
Start
Adapting the
dominant
Prepare the injector
32
/ 52
If the LED in the start key flashes,
◆ press the key.
– The compression device moves into park position, if present
– Oblique setting moves into the 0° position
Once these positions are reached, the flashing LED goes out.
◆ Select an organ program for Periscanning
◆ Position the patient and set the following parameters with the corresponding
operating elements:
– Primary collimation (rectangular diaphragm)
– Semitransparent wedge filters (if option is available)
– Cu prefilter (preferably filed in the organ program)
Position the patient so that the region the vascular filling of which should be im-
aged first is located in the central ray.
◆ Press one/several keys,
– The dominant preset in the organ program is overwritten.
If a left / right dominant or both dominants are selected, the system switches
over automatically to the central circular dominant if the preselected dominant is
covered by the filters, since this would inevitably result in a faulty exposure.
◆ Connect the injector plug to the electronics cabinet (only with radiation-syn-
chronous injector release (option)).
◆ Set the injection data and switch to ready for injections.

Test run

The X-ray system stands in the starting position for the scanning process.
Start a test run without radiation.
◆ Switch the light localizer of the collimator on.
◆ Deflect the joystick up / down or the key to the left / right until the required
examination area has been scanned.
◆ If necessary correct the starting position.
AXD3-340.620.13.01.02
Operator Manual