Siemens AXIOM Iconos R200 Operator's Manual page 183

Table of Contents

Advertisement

Adjustment
Removal
Maintenance note
AXIOM Iconos R200
Warning
Be sure to check that the footboard has locked into position before allowing pa-
tients to sit or stand on it. If you want to tilt the patient table from the horizontal
position to the Trendelenburg position, you must secure the patient with the
shoulder supports or the foot restraints at 15
tion >40°, the foot restraints must be used in addition to the shoulder supports.
Make sure that patients standing on the footboard hold onto the handgrip and
the handgrip strip. For additional safety, use the compression belt for frail or
very unsteady patients.
◆ Press the spring-loaded handgrip strips in the handgrips;
– the two red buttons must be visible;
◆ Move the footboard to the desired position (lock-in position);
◆ Release the handgrip strips and shift the footboard slightly back and forth until
it locks in;
– the two red buttons must be flush with the handgrip;
Warning
The footboard may only be unlocked, adjusted or removed when the table is in
the horizontal position. In all other system positions, the footboard could fall off
and present a high risk of accidents.
◆ Move the patient table to the horizontal position;
◆ Press the spring-loaded handgrip strips in the handgrips;
– the two red buttons must be visibible;
◆ Lift the footboard first on one side and remove from the accessory rail;
◆ Remove other side of the footboard from the accessory rail.
In order to prevent influx of contrast medium into the sockets, the pedal must be
covered with absorbent fleece material.
AXD3-340.620.13.01.02
Accessories
Standard Accessories
o
. In case of a Trendelenburg posi-
11
/ 54