U1 Power Supply Unit 5 V ± 5% / 15 V ± 10 - Siemens POLYMOBIL Plus Service Instructions Manual

Hide thumbs Also See for POLYMOBIL Plus:
Table of Contents

Advertisement

2 - 6
Measurement points:
VP
P12
GNDL
X4
U1 power supply unit 5 V ± 5% / 15 V ± 10 %
Fuse: 2 A T
U2 power supply unit 12.5 V ± 10 %
Fuse: 5 A T
POLYMOBIL Plus
±
+15 V
10 % for inverter frequency
+ 12.5 V
Ground (+ 12.5 V)
Pin
Color
1
red
2
red
3
red
4
black
5
black
6
black
Checking the assemblies
±
10 % auxiliary voltage
SPR8-125.061.02
Rev. 01
01.00
±
+12.5 V
10 %
±
+12.5 V
10 %
±
+12.5 V
10 %
Ground (+12.5 V)
Ground (+12.5 V)
Ground (+12.5 V)