Releasing fluoroscopy

❏ Use the foot switch to release fluoroscopy.
❏ Radiation is released as long as you continue to press the footswitch for fluo-
roscopy. After you let the switch go, the last image is retained and displayed
on the monitor (LIH).
❏ With the red emergency STOP button actuated, fluoroscopy can be switched
back on consciously after letting the fluoroscopy switch go and pressing it
again.
❏ A patient folder must be created and selected in the FLUOROSPOT Compact.
❏ No processing functions apart from storing the image are possible during flu-
oroscopy.
❏ The radiation ON indicators must light up.
Warning
If these radiation ON indicators light up although no switch has been actuated
and the current fluoroscopic image cannot be seen on the monitor:
◆ Switch the system off immediately!
◆ Notify the SIEMENS Uptime Service.
◆ Discontinue use of the system.

Fluoroscopic data

The fluoroscopic data of kV, mA, time and area dose product (optional) are dis-
played on the integrated generator control console.

Display of the kV and mA values

❏ During fluoroscopy:
– Running display of the momentary kV and mA values
❏ After the end of fluoroscopy:
– kV and mA mean value of the examination

Display of the fluoroscopic time

Continuous display of the time accumulating in the examination in minutes.

Display of the area dose product (optional)

Continuous display of the area dose product accumulating in the examination.