Dr-Scanning (Optional); Organ Programs; Creation Of Programs; Temporary Changes Of Programs - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

Examination
Digital Radiography
40
/ 52

Organ programs

All parameters necessary for the exposure are stored in organ programs.
➩ Refer to the FLUOROSPOT Compact Operator Manual and the POLYDOROS
SX register for further details on the organ programs and their programming.

Creation of programs

➩ Refer to the FLUOROSPOT Compact Operator Manual.

Temporary changes of programs

For the following parameters stored in the organ program, it is possible to change
them temporarily with the corresponding operating elements:
– Dominants
The changed parameters apply only until you select another / same organ pro-
gram or switch the system off.

DR-Scanning (optional)

General
To display large objects, a series of displayed images is combined to give a large-
format image in the image processor of the workstation.
The entire exposure area is moved over continuously with the image receptor
proceeding from the starting position (not in Peristepping).
After the acquisition the images are sent from the imaging system to the LE-
ONARDO workstation for processing (e.g. image reconstruction).
Depending on the examination, there are 2 types of acquisition:
❏ DR scanning for
– images of the spine
– orthopedic longleg images
– abdominal survey images
❏ Peristepping for survey images of the peripheral vessels
Generate an organ program for the scanning technique in the series image
menu.
AXD3-340.620.13.01.02