Selecting Direct Exposure Technique; Select 0-Point Technique - Siemens AXIOM Iconos R200 Operator's Manual

Table of Contents

Advertisement

Selection
Deselection
Exposure data
to be set
Display of the
exposure data
AXIOM Iconos R200

Selecting direct exposure technique

Select 0-point technique

❏ Setting the kV value from fluoroscopy
❏ Exposure cut-off by the automatic exposure control.
Iconos DFR
Iconos
Iconos
Cassette
Cassette
45
kV
0.3
mA
0.0
0.1
min
0
%hu
H
3.4
µGy*cm 2
◆ Select an organ program with 0-point technique
◆ Actuate
– the mAs button: 2-point technique is selected or
– the ms button: 3-point technique is selected
With organ program:
◆ Select an organ program without 0-point technique
❏ The following exposure data can be set:
– Focal spot
– Tube load
– Measuring fields (dominants)
– Film-screen combination
– Density compensation
❏ After the end of the exposure, the mAs and ms values in the display are flash-
ing.
AXD3-340.620.13.01.02
Iconos Bed
Vertix
40
kV
U
D
80%
100%
100%
POLYDOROS SX 65/80
Integrated Control Console
Info
mAs
ms
Program
0
0
21
/ 36