2.13

DAP measuring system (optional)

Press the test button on the display. The value on the display should be in the range
80-120 µGym
high sensitivity, i.e., DAP resolution 0.01 µGym
to the Service Instructions. (when using a chamber of high sensitivity, i.e., DAP resolu-
tion 0.01 µGym
X
Check the function of the DAP measuring system by releasing an exposure. Check if a
measured value is shown on the display.
SIE
Function of DAP measuring system
POLYMOBIL Plus
2
(DAP resolution 0.1 µGym
2
). If not, calibrate the system according to the Service Instructions.
SPR8-125.831.01.02.02
05.05
Inspection and Maintenance
2
2
) or 8-12 µGym
2
). If not, calibrate the system according
Page 38 of 40
CS PS 24
(when using a chamber of
Siemens AG
Medical Solutions