Remote Exposure Switch System (Optional); Dap Measuring System (Optional); Test Of Kv-Accuracy - Siemens MOBILETT Plus HP Installation And Setting Instructions Manual

Table of Contents

Advertisement

Installation and setting

Remote exposure switch system (optional)

Check the function of the remote exposure switch system according to chapter functional
check of "Installation Instructions: Installation of remote exposure switch system".

DAP measuring system (optional)

Check the function of the DAP system according to section "Monthly checks" of the
Instructions for use.

Test of kV-accuracy

Requirement:
- The measured kV values, shall be within the limits.
Required measurement devices:
- kV-meter, based on the filter-comparison method, for example PTW-Nomex kV meter.
When measuring the kV non invasive a correction factor from the user manual of the
kV-meter has to be used. If a kV meter is not available and if it is within local laws and
regulations, the measurement can be made with an oscilloscope on D1. KVPEAK - D1
GND with a result of 40 kV/V.
NOTE
For the actual filtration of this MOBILETT, see the test protocol.
If the actual inherent filtration is unknown, the correction shall be made with a reference
value of 5 mm Al.
Put the kV meter into operation, see operating instructions for the kV-meter.
Place the measurement detector on a table or other. Using the light field, set the detector
in the beam projection, see Fig. 5.
Set the SID to approx. 50 cm (20"). The SID may vary between different kV-meters.
Fig. 5
Siemens-Elema AB
Solna, Sweden
Higher filtration results in higher kV values.
If the MOBILETT is equipped with a DAP measuring system;
Remove the DAP ionization chamber first.
SPR8-220.033.03
Rev. 02
Page 7 of 12
10.02
SPS-UD
2 - 7
2
2
2
MOBILETT Plus HP