General

Safety information - risk of infection

WARNING

Information on the protective conductor resistance test

Observe the instructions in the "Safety Rules for Installation and Repair"
(ARTD-002.731.17 ...).
The protective conductor resistance must be measured after every intervention in the sys-
tem.
However, documentation of the measured values is required only during periodic safety
checks.
If parts/components that can significantly influence the protective conductor resistance
(e.g., replacement of the power cable, replacement of the power-up module, replacement
of multi-pole connection cables which also create the protective conductor connection
between system parts (e.g., monitor cable or C-arm cable)) are replaced or if protective
conductor connections have been repaired, the protective conductor resistance must be
measured. The values must be documented and evaluated in the protective conductor
resistance protocol.
Siemens
Medical Solutions
Risk of infection due to pathogens!
Non-compliance can lead to severe injury and even death.
This product can be contaminated with infected blood
or other bodily fluids.
Avoid all contact with blood or other bodily fluids!
Strictly observe the safety information in
ARTD-002.731.37.. regarding prevention of infectious
diseases during customer service calls.