PC Board Replacement

Adjustment test

Set DIP switch S2.4 to "OFF".
Close the "height" and "width" collimator blades completely by hand.
Remove the prefilter (if present).
Move the iris blades to an opening of 70 x 70 mm (measure at the iris blades).
Switch the system back on or reinsert fuse F1 on D1.
Switch DIP switch S2.4 to "ON" for initialization.
The system will initialize. DIP switch S2.4 should remain "ON" during operation.
The height and width blades will open or close to the 44mm format (± 0.75%) x 44 mm
(± 0.75%) at a SID of 125 cm (35 mm x 35 mm at 100 cm SID).

Control position of the collimator blades

Install the cassette holder. Do not insert a film cassette.
Move the system into the mid-range position.
Release fluoroscopy.
A radiation field of comparable size
44 mm (± 0.75%) x 44 mm (± 0.75%) - is visible on the monitor at a SID of 125 cm.