5.
FUNCTIONAL VERIFICATION
5.1.
Introduction
Functional
verification
should
be
per-
formed annually during scheduled service
intervals,
whenever
proper
operation
of
the
monitor
is suspect,
and
after the
monitor has been closed following recali-
bration
or
repair
These
procedures
validate proper operation of the monitor
hardware.
Many
of the monitor's
func-
tions, such as alarms and waveforms and
scale sizing, are software operations. The
monitors
software
is
automatically
checked during the monitor power-up and
on-line selftests.
All software
functions
are performing in accordance with design
specifications
if no
error
messages
are
displayed. Some error messages relate to
hardware failures.
This chapter describes the test methods
used to verify the functional operation of
SIEMENS SIRECUST 1280 and 1281
bedside monitors. When
performing the
following tests, always use the specified.
recommended
equipment
or
a known
equivalent.
Use
of
unauthorized
test
equipment
could
result
in
inconclusive
tests or damage to the monitor.
If the monitor should need calibration or
fail to perform according to specifications,
the
procedures
in
this
chapter
refer
service personnel to other chapters in this
Manual, for recommended service.
5.2.
Tools and Test Equipment
The
tools
and
test
equipment
recom-
mended
for
the
functional
verification
procedures
are
listed
in
Table
5-1.
Substitutions
are
approved
only
if an
equivalent is listed.
All
test
equipment
specified
in these
procedures is assumed to be operating in
accordance
with
the
manufacturers
specifications.
Table 5-1.
Recommended Tools and Test Equipment for Functional Verification
D
UNIT
SIREM Remote Cartridge Unit
SIRECUST LIM Cartridge
60 Hz.
50 Hz.
50 Hz. PTB
SIREM
Cable
5-Lead EKG
Patient Cable
5-Lead Set
SDL 4 Cable
DESCRIPTION
SIEMENS Art. No.
SIEMENS Art. No.
SIEMENS Art. No.
SIEMENS Art. No.
SIEMENS Art. No.
or Art. No.
SIEMENS Art. No.
SIEMENS Art. No.
SIEMENS Art. No.
SIEMENS Art. No.
85 39 926 E2506
89 92 020 E2507
88 62 252 E2507
88 61 700 E2507
89 44 787 E2513
87 93 440 £2513
45 27 529 EH405
45 28 030 EH405
87 94 810 E2513
84 28 450 RE999.
12" Video Alignment Grid
SDL Breakout Box
Temperature Simulator Test Plug
Cardiac Output Test Plug
Digital Multi Meter (DMM), 4% digit
Leakage Current Analyzer
Watt Meter (Optional)
Polymed Simulator
or
Waveform Simulator
Ruler, Precision Millimeter Grade
SIEMENS Art. No.
12 66 811 EE50U
See Appendix E. Buildable Test Eguipment
See Appendix E. Buildable Test Eguipment
Fluke, Model 8050A (or eguivalent)
.
Dynatech-Nevada, Model 432D (or equivalent)
Valhalla Scientific, Model 2101 (or equivalent)
SIEMENS Art. No.
45 28 345 EH436
Dynatech-Nevada, Model 215A (or equivalent)
Page5-1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sirecust 1281