17 Troubleshooting

DANGER!
Danger to life due to electric shock!
If a fault occurs ensure the electric motor, power box and other electrical parts checked by qualified personnel
only.
Do not open protective covers of the electric motor or power box.
Error/Fault
Adjusting with position buttons not
possible
Faulty mattress platform height/tilt adjus-
tment
Lowering backrest from the upright position
not possible
Adjusting siderails not possible
Faulty brakes
Mattress not inflating
Unable to change mode or pressure
Fault symbol illuminated & audio alarm
CPR symbol illuminated
66
Cause
GO button was not pressed
Function disabled on Attendant Control Panel
Drive motors have no power,
Defective drive motors,
Defective battery
Plug inserted incorrectly
Faulty power source.
Faulty control element
There is an object on the undercarriage cover
Function disabled on Attendant Control Panel
Drive motors have no power,
Defective drive motors,
Defective battery
Plug inserted incorrectly
Faulty power source
Faulty control element
There is an object under the back- rest or in the
drive mechanism
Locking handle is defective
The siderail lock is dirty
Locking handle is defective
Dirt blocking brakes mechanically
The brake mechanism is defective
SCU mains switch turned off
No power to bedframe
Mains plug inserted incorrectly
Faulty power source
Faulty SCU
Air leaking
CPR valve leaking or open
Air pipes blocked
Mattress partialy inflated
Mattress damaged or faulty
Go button was not pressed
One time exception fault
Persistent reoccuring fault
In CPR mode
Solution
Press the GO button.
Enable disabled function.
Check the mains connection.
Notify the service department.
Insert the mains plug correctly.
Notify the service department.
Notify the service department.
Remove the object.
Enable disabled function.
Check the mains connection.
Notify the service department.
Insert the mains plug correctly.
Notify the service department.
Notify the service department.
Remove the object.
Notify the service department.
Clean the locking mechanism.
Notify the service department.
Clean the brake system.
Notify the service department.
Turn switch on. Green mains
power switch will illuminate.
Check mains connection.
Insert mains plug correctly.
Notify the service department
of the manufacturer.
Notify the service department
of the manufacturer.
Check mattress pipe connec-
tion.
Check CPR valves closed.
Check piper are not trapped or
kinked.
Check mattress is unfolded
and flat.
Notify the service department
of the manufacturer.
Press the GO button.
Switch off SCU & switch back
on, reset to see if failt is self-
-cleared.
Notify service department,
mute audio alarm.
Press CLP or Max to cancel
D9U001MC5-0101_08