12.1.2 Taring
The taring is done in the range from 5kg to 249,5kg. The taring is used to set "0" on the display before placing the patient on the
bed. It is used to show actual weight of the patient.
The taring must be done on the unloaded bed without patient. It is recommended to position mattress platform about 20 cm above
the lowest position and the mattress platform in the horizontal position.
To tare weight:
►
Ensure that nothing touches the bed except you.
►
Press icon 2 (Zero/T) for 0,5 s until the display starts to flash.
►
Press icon 2 to confirm taring. The „0" is shown on the display.
Place the patient on the bed.
To cancel taring:
►
Press icon 5 while taring.
12.1.3 Displaying
Verification Scale Interval is 0.5 kg.
►
Press button 1 to display value with actual scale interval 0,1 kg for 5 s.
Display shows normally actual weight if other functions are not activated.
12.1.4 Hold Mode
Hold Mode must be used only when the scales are stabilized.
It allows adding or removing bed accessories and other items without changing the weight value.
To activate Hold Mode:
►
Wait 5 s until the scales are stabilized. The LED 4 will be illuminated when the scales are stabilized.
►
Press button 6 for 2 s.
►
The display shows „HoLd", indicator of activated Hold Mode is illuminated.
►
Add or remove required accessories.
To deactivate Hold Mode:
►
After adding or removing accessories wait 5 s, until the scales are stabilized on the display.
The LED 4 will be illuminated when the scales are stabilized.
►
Press button 6 for 2 s.
►
The display shows the original weight value and the indicator of Hold Mode is not illuminated.
To deactivate Hold Mode without fixing the weight value:
►
Press button 5.
12.1.5 Setting Mode
To set date, date format, time and unit of weight:
►
Press button 2 and button 6 simultaneously for 3 s.
The value to be changed flashes on the display.
To navigate in list:
►
Press button 7 to shift to the next parameter:
1.
minutes
2.
hours
3.
date format (month-day/day-month)
4.
year
5.
month
6.
day
7.
unit of weight
To set the corresponding value:
►
Press button 2 or button 6 to increase or reduce the value.
To leave setting mode:
►
Press button 5.
Setting Mode is left without saving the last setting.
D9U001MC5-0101_08
45
Need help?
Do you have a question about the Multicare LE and is the answer not in the manual?
Questions and answers