User Manual and Technical Description ELEGANZA 5 Positionable Bed for Intensive Care and OptiCare integrated mattress replacement system D9U001GE5-0110 Version: 06 Publication Date: 2018-10...
19 Warranty ..............92 20 Standards and Regulations ........93 20.1 Eleganza 5 ..............93 20.2 OptiCare ..............93 20.3 Manufacturer .............93 1 Symbols and Definitions 1.1 Warning Notices 1.1.1 Types of Warning Notices Warning notices are differentiated by the type of danger using the following key words: ►...
Page 5
SAFE WORKING LOAD PHYSICAL DESCRIPTION OF AN ADULT (DESIGNATION OF MEDICAL BED FOR ADULTS) USE MATTRESS RECOMMENDED BY MANUFACTURER DO NOT PUT ANY OBJECTS ON UNDERCARRIAGE CALFREST LOAD LIMIT WARNING AGAINST CRUSHING OR TRAPPING OF HANDS JACK FOR ATTACHMENT OF CONDUCTOR FOR POTENTIAL EQUALISATION CPR LEVER GENERAL WARNING SIGN CAUTION...
Page 6
SAFETY ISOLATING TRANSFORMER (GENERAL) FOR INDOOR USE ONLY MAXIMUM PATIENT WEIGHT WEIGHT OF BED MET MARK FOR Eleganza 5 CE MARK FOR OptiCare WEEE SYMBOL (RECYCLE AS ELECTRONIC WASTE, DO NOT PUT INTO THE HOUSEHOLD WASTE) RECYCLING SYMBOL RECYCLING SYMBOL...
MANUFACTURER MANUFACTURING DATE REFERENCE NUMBER (PRODUCT TYPE DEPENDING ON CONFIGURATION) SERIAL NUMBER 1.5 Symbols and Labels on the Mattress DO NOT IRON! DO NOT USE PHENOL! DO NOT WRING! REGULARLY INSPECT THE INSIDE OF THE COVER FOR CONTAMINATION MACHINE WASH AT MAX. 71°C FOR 3 MINUTES D9U001GE5-0110_06...
Page 8
TUMBLE DRY ON LOW HEAT SETTING (MAX. 60°C) HANDWASH WITH DETERGENT (INITIAL TEMPERATURE OF HOT WATER SHOULD NOT EXCEED 50°C) DISINFECT USING SOLUTION CONTAINING LESS THAN 1000 ppm OF CHLORINE (REFER TO INSTRUCTIONS FOR USE) WIPE WITH WATER COVER MATERIALS ARE FIRE RESISTANT TO BS7175, SOURCE 0, 1 AND 5 MATTRESS FOOT PART (OptiCare) D9U001GE5-0110_06...
1.6 Serial Labels with UDI Fig. Serial Label with UDI (Eleganza 5 with scales) Address of Distributor Manufacturing Date (Year-Month-Day) DI (Device Identifier) / GTIN (Global Trade Item Number) 1D Bar code GS1-128 (Serial Number) Symbols Configuration number Electrical Specification...
Page 10
Fig. Serial Label with UDI (OptiCare - SCU) Fig. Scales label (WS17) Fig. Serial Label with UDI (OptiCare - mattress) Address of Distributor Manufacturing Date (Year-Month-Day) DI (Device Identifier) / GTIN (Global Trade Item Number) 1D Bar code GS1-128 (Serial Number) Serial Number PI (Product Identifier) 2D Bar Code (GS1 DataMatrix) DI+PI=UDI...
1.9 Definitions Basic Bed Configuration the pricelist model configuration, not including a mattress Bed Weight The value depends on the product configuration, accessories or custo- mer adjustments. Clearance of Undercarriage the height from the floor to the lowest point of the undercarriage between the castors, for the manipulation of accessories under a braked bed in the standard position Duty Cycle...
2 Safety Instructions WARNING! Eleganza 5 bed should be left in its lowest position when the patient is unattended in order to reduce risk of injury due to falls! WARNING! Siderails of Eleganza 5 should be located in the „up“ position to reduce the risk of the patient accidentally...
Page 14
WARNING! During specific investigations or treatments the significant risks of reciprocal interference posed by ME equi- pment may occur. ► Follow the instructions carefully. ► Use the bed exclusively if it is in perfect working order. ► If necessary, check the bed functions daily or at each shift change. ► Ensure any user has read and understood this manual completely before operating the product. ►...
3 Intended Use Eleganza 5 is a positionable bed for intensive care. Its purpose is to support patient and to facilitate treatment and manipulation with patient for nursing personnel. Eleganza 5 is equipped with weighing system designed for weighing pa- tients.
4 Product Description 11 10 Fig. Bed Overview (Eleganza 5) Removable Foot Board Split Siderail – Middle Siderail with Integrated Control Panels for Patient Split Siderail – Head Siderail with Integrated Control Panels for Patient Removable Head Board iBoard Standard CPR Lever –...
5 Technical Specification All technical data are rated data and are subject to construction and manufacturing tolerances. WARNING! If Eleganza 5 bed is used with OptiCare integrated mattress system respect values of mechanical and electrical specifications which can harm none of them! 5.1 Identification of Applied Parts (Type B) All part of the bed (and accessories) the patient can reach are type B Applied Parts.
Pb AKU 2 x 12 V / 1,2 Ah / Fuse 15A Fuse 2 x T2.0A L 250V for 230 V NOTE Upon request, Linet ® can deliver hospital beds with electrical specifications that comply with regional standards (custom voltage, different mains plugs). 5.6 Mechanical Specifications (OptiCare) Parameter...
5.8 Electrical Specifications (OptiCare) Input Voltage, Frequency 100/127 VAC, 50/60 Hz (Model 110V) Maximum Power Input max. 40 VA (when operating from mains supply) Ingress Protection (EN 60529) IPX4 Protection Class Class I with Applied Parts Type B Electrical Safety in conformity with EN 60601-1 Fuse 2 x T1A Anti-surge Fuse 5.9 Electromagnetic Compatibility...
► Use exclusively Hospital Grade or Hospital Only receptacles for grounding. Eleganza 5 and OptiCare are designed for use in rooms for medical purposes. Electrical installations must therefore meet local norms laying down the necessary conditions for electrical installations. ►...
8 Initial Instructions Set up the bed as follows: ► Unpack the bed. ► Check the delivery (see Scope of Delivery and Bed Variants). ► Remove isolating foil from the mains control box (see Battery Activation). ► Install equipment and accessories (see Installation). ►...
Slide bed ends into sleeve fitting. ► Lock sleeve fitting. Fig. Locking of the foot board Mattress Platform Eleganza 5 without x-ray cassette holder has four removable parts of Mattress Platform (1,2,3 and 4). LEFT SIDE RIGHT SIDE Fig. Parts of Mattress Platform D9U001GE5-0110_06...
8.2.1 Potential Equalisation The bed is equipped with a standard protective connector. This connector is used for potential equalisation between the bed and any intravascular or intracardiac device connected to the patient to protect the patient from static electric hocks. Fig.
Unsuitable surfaces: ■ Too soft, unsealed or defective flooring ■ Soft wooden flooring ■ Soft and porous stone floors ■ Carpeted floors with underlay ■ Soft linoleum ► For longer distances, ensure that the castor steering function (main control) is activated. ►...
Signalisation The LED (on iBoard Standard or Additional Supervisor Panel) indicates the battery’s charge status: Yellow LED Battery charge status Not lit Battery capacity is sufficient (charging completed) Short flashing (shortly lit, longer not lit) (circa 1.8 sec.) Battery is charging - continue charging until the LED is extingu- ished.
To cancel this status: ► Press button 10.2 Removing the Bed from Service Remove the bed from use as follows: ► Disconnect the bed from the mains. ► Disconnect the ground wire. ► Deactivate the battery. ► Remove accessories. To prevent damage during storage: ►...
POSITIONING iBoard Additional Handset Integrated Control Foot Control- Foot Control- Standard Supervisor Panels (head side- Height Tilt Panel rail, foot siderail) Backrest Thighrest Calfrest Bed Height Autocontour Examination Position Emergency Trendelenburg Position Anti-Trendelenburg and Trendelenburg Tilt Bed Extension CPR Position Cardiac Chair Position Mobilization Position Lateral Tilt...
Page 29
1. Display 2. Keyboard - Scales Section 3. Keyboard - Bed Exit/BedMonitor Section 4. Keyboard - Positioning Section 5. Keyboard - Settings Section or Setting Section and Integrated Mattress Section Fig. iBoard Standard with Integrated Mattress Keyboard Settings Section 1 2 3 4 Fig.
Page 30
Available options are: YEAR / MONTH / DAY / HOUR / MINUTE / WEIGHT TIMER. It is not possible to set value for option WEIGHT TIMER in the version of Eleganza 5 bed without scales. NOTE: Options are displayed in this order. YEAR follows after WEIGHT TIMER again.
Page 31
► To set the value use buttons ► Press button to save the value and continue to the setting of the other option. The value is saved by exiting. Icon 5 indicates another option automatically and icon 8 indicates its actual value. To set day: ►...
Page 32
Positioning Section 22 21 20 15 16 Fig. Positioning Section - Display and Keyboard 1. Autocontour Adjustment Buttons (simultaneous movement of the Backrest and Thighrest) 2. Battery LED 3. Mains Power LED 4. Anti-Trendelenburg Tilt Button 5. Locked Backrest LED 6.
Page 33
GO BUTTON The button activates the keyboard of all Control Elements. Pressing button will keep the keyboard active for 10 minutes. Pressing a button will keep the keyboard active for another 10 minutes. During this time the following is possible: ►...
Page 34
Pop-ups (iBoard Standard) Pop-up Meaning Required Action LOCK Function locked Unlock function X-RAY Incorrectly inserted X-Ray Cassette Insert X-Ray Cassette Holder Holder correctly GO Button not activated Activate GO Button SIDERAIL Siderail folded down (Lateral Tilt disa- Raise siderail up to enable Late- bled) ral Tilt Prevention from collision with floor during...
11.1.2 Additional Supervisor Panel The Additional Supervisor Panel is a standard Control Element. The Additional Supervisor Panel can be hung on the foot end or on siderails if required. It is possible to hold the Additional Supervisor Panel in the hand while operating. ►...
Function Buttons The function buttons 4, 5, 7, 8, 10, 11, 12, 15, 16 and 17 are described in chapter Bed Positioning. NOTE Pressing two function buttons at the same time will be recognized as an error by the controller. The controller will interrupt immediately all bed movements and display shows alert.
1 2 3 1. Thighrest Adjustment Button 2. Thighrest/Backrest Lock LED 3. Backrest Adjustment Button 4. GO Button 5. Autocontour Adjustment Button 6. Flashlight Button 7. Height Lock LED 8. Bed Height Adjustment Button The function buttons 1, 3, 5 and 8 are described in chapter Bed Positioning. To switch on the flashlight: ►...
11.1.5 Foot Control Bed Height (optional) The foot control is optional and allows setting the Height of the bed with one’s feet. 1. Protection Frame against Unwanted Activation 2. Raise Mattress support platform Pedal 3. Examination Position Pedal 4. Lower Mattress support platform Pedal Fig.
11.2 Bed Positioning 11.2.1 Backrest To position Backrest use: ► iBoard Standard ► Additional Supervisor Panel ► Handset ► Integrated Control Panel for Patient (in middle siderail) ► Integrated Control Panel for Patient (in head siderail) Fig. Backrest Angle on iBoard Standard Display iBoard Standard Display shows Backrest Angle.
11.2.2 Thighrest To position Thighrest use: ► iBoard Standard ► Additional Supervisor Panel ► Handset ► Integrated Control Panel for Patient (in middle siderail) ► Integrated Control Panel for Patient (in head siderail) iBoard Standard: ► Press button ► Press selected part of Thighrest Adjustment Button until intended position is reached.
11.2.3 Calfrest To position Calfrest position Thighrest firstly. To position Calfrest use: ► iBoard Standard ► Additional Supervisor Panel iBoard Standard: ► Press Button ► Press selected part of Calfrest Adjustment Button until intended position is reached. Fig. Calfrest Adjustment Button (iBoard Standard) 1. Calfrest Up 2.
Page 42
iBoard Standard: ► Press Button ► Press selected part of Bed Height Adjustment Button until intended position is reached. Fig. Bed Height Adjustment Button (iBoard Standard) 1. Mattress Platform Additional Supervisor Panel: ► Press Button 2. Matress Platform ► Press selected part of Bed Height Adjustment Button Down until intended position is reached.
11.2.5 Autocontour To position Autocontour use: ► iBoard Standard ► Handset ► Integrated Control Panel for Patient (in middle siderail) ► Integrated Control Panel for Patient (in head siderail) iBoard Standard: ► Press Button ► Press selected part of Autocontour Adjustment Button until intended position is reached.
11.2.6 Emergency Trendelenburg Posi- tion Trendelenburg position provides anti-shock conditions for the patient. During Trendelenburg Position Mattress Plat- form is straightened in the tilt. To position Emergency Trendelenburg Position use: ► Additional Supervisor Panel Fig. Trendelenburg Position Additional Supervisor Panel: ►...
11.2.9 Bed Extension To position Bed Extension use: ► Additional Supervisor Panel Additional Supervisor Panel: ► Press Button ► Press selected part of Bed Extension Adjustment Button until intended position is reached. Fig. Bed Extension Adjustment Button (Additional Supervisor Panel) 1.
11.2.11 Cardiac Chair Position To position Cardiac Chair Position use: ► iBoard Standard ► Additional Supervisor Panel iBoard Standard: ► Press Button ► Press Cardiac Chair Position Button until intended position is reached. Fig. Cardiac Chair Position Button (iBoard Standard) Additional Supervisor Panel: ►...
11.2.13 Lateral Tilt To position Lateral Tilt use: ► iBoard Standard ► Foot Control Tilt ► Additional Supervisor Panel Fig. Lateral Tilt Angle on iBoard Standard Display iBoard Standard Display shows Lateral Tilt Angle. Maximum Lateral Tilt Angle is 15 degrees. iBoard Standard: ►...
11.3 Scales Control Use iBoard Standard to control the scales. iBoard Standard Fig. Scales Section - Display and Keyboard 1. ZERO/T Button (tare or zero scales) 2. WEIGHT/CLEAR Button (cancel) 3. HOLD Button 4. Memory Button (save or display weight value) 5.
Page 50
► Press button again to confirm taring. “0” is shown on the display. Place the patient on the bed. To cancel taring: ► Press button while taring. 3) Displaying Verification Scale Interval is 0.5 kg. Field 6 shows normally actual weight value. NOTE Weight value automatically disappears after 1 minute.
8) Zeroing Scales Zeroing is only possible in a range of ±5kg from factory zero. Zeroing is used to reset weight on the display and set up user zero, which sets the maximum weight range of the weighing system. Zeroing must be done with an empty, unloaded bed, without the mattress and accessories. Zeroing is done after installation, weight verification or servicing.
Page 52
1) Preparation ► Place a patient on the bed with suitable mattress. NOTE For the correct Bed Exit Monitoring in the Inner Zone patient´s position in the middle of the bed is needed. 2) Activation of Bed Exit Monitoring Bed Exit Monitoring is OFF and icon is displayed by default.
Page 53
5) PAUSE During PAUSE Mode Bed Exit Monitoring is temporarily interrupted and Alarms are not activated. To PAUSE Bed Exit Monitoring: ► Press button Icon 7 appears on the display with 15 minute countdown timer. After PAUSE period elapsed and patient is in ordered position Bed Exit Monitoring is reactivated. To extend the PAUSE period: ►...
11.5 CPR Backrest Release WARNING! Risk of injury due to lowering the backrest too quickly! ► Ensure that the siderails are in the low position. ► Ensure that there are no body parts between any movable parts of the bed. ►...
11.7 Siderails The split siderails are components of the bed in contact with patient. A pneumatic spring supports the operation of the split siderails. The nursing personnel are responsible for the siderails being raised up while the patient is in bed. WARNING! Risk of injury, damaging or unintentional movement of the bed due to incorrect placement of accessories or Handset!
To release siderails down: ► Press Siderail Release Handle (3) inwards. ► Unlock siderail by pulling Siderail Release Handle. ► Fold down siderail slowly. 12 Equipment 12.1 i-Brake (optional) ® It is possible to equip the bed with an automatic castor brake. The automatic castor brake prevents injuries of patients and staff due to an unbraked bed.
Page 57
Each bed can transport only single patient at a time and cannot be used to transport other items (except bed accessories in secured position). NOTE For information concerning uses other than those outlined in the “Specifications of Use” section above, please contact Linet ®. Manipulation CAUTION! Damage to i-Drive Power main...
5. ON Button (i-Drive Wheel Activa- tion Button) 6. OFF Button (i-Drive Wheel Retraction and Deactivation Button) 7. FAST FORWARD Button 8. STOP DRIVE Button 9. ON LED 10. Mains Power LED (bed connec- Fig. Activation Panel (4) ted to the mains or bed braked) 11.
Fig. i-Drive Power Mains Switch with Label 1. i-Drive Power Mains Switch 2. i-Drive Power Emergency Retraction Button Fig. i-Drive Power Emergency Retraction Button Label 12.3.2 Powered Drive CAUTION! Damage to property due to incorrect transport and involuntary movement! ► Prior to transport, ensure that the bed is disconnected from the mains.
NOTE Always brake the bed by using the castor control lever when the transport is finished or interrupted. The i-Drive electromagnetic brake is not designed to permanently brake the bed. NOTE In a crisis situation (e.g. acceleration when driving down a steep slope) i-Drive dual braking prevents acceleration and slows down bed movement.
12.6 X-Ray Lung Examination (optional) The Backrest of the bed consists optionally of HPL and is x-ray translucent. The bed is equipped with an x-ray cassette holder inse- ted under the Backrest left side. This design allows taking x-ray images of the patient’s lungs without moving the patient manually. Necessary Steps before the Examination NOTE This procedure is above all suitable for patients...
12.7 Nurse Call Button for activating the Nurse Call function: The buttons for activating the Nurse Call function are located on the inner and outer sides of the head siderails. Speakers and microphones are located on the inner sides of the head siderails. Fig.
CAUTION! Incompatibility with bed due to incorrect mattress dimensions! ► Check maximum approved mattress dimensions (chapter Technical Specification). The manufacturer recommends the use of the following mattress systems on the Eleganza 5 bed: PASSIVE MATTRESSES ACTIVE MATTRESSES ■ CliniCare 10 ■ ...
13.3 OptiCare (integrated mattress) 13.3.1 Intended Use OptiCare is integrated mattress replacement system for use with Eleganza 5 hospital bed. Eleganza 5 with OptiCare suitable for: ► patients at moderate risk levels ► patients with any stage/category of pressure ulcer 13.3.2 Contraindications...
Page 66
The heel cells will self inflate when the bed is lengthened. Safety Straps The sides of the foam base fit the sides of the Eleganza 5 Mattress Platform to prevent the mattress from shifting when the patient is getting into or out of bed.
SCU (System Control Unit) The SCU (System Control Unit) is installed under the leg section of the bed frame. It is possible to have the Eleganza 5 bed delive- red with the SCU factory-fitted, or to have it retrofitted by approved service personnel.
Standard control screen. If replacing the OptiCare mattress with one that is not from the Linet integrated systems range then you will need to cancel the Mattress Not Connected alarm.
13.3.6 Preparing OptiCare for patient DANGER! Risk of suffocation due to air-impermeable mattress cover! ► Use mattress cover correctly. ► The nursing staff are responsible for the safety of the patient on the mattress cover. WARNING! Risk of injury when positioning patient on the bed! Before positioning the patient on the bed: ►...
Page 70
To mute compressor of the integrated mattress: ► Press button Icon indicates activated Mute Mode. OptiCare Mattress Controls Control and information on status of the OptiCare mattress is by the diplay and keyboard on the iBoard Standard. Patient in bed detection (PIB) Patient in bed detection system detects when a patient has entered or left the bed.
Page 71
MATTRESS PREPARED FOR PATIENT When „MATTRESS INFLATION“ text disappears and Mattress icon is fully green the mattress is ready for a patient. NOTE Flashing Mattress Icon with selected OPT Mode indicates continuing inflation. To use the mattress: ► position patient on the mattress. Fig.
Page 72
PRESSURE COMFORT ADJUSTMENT The mattress pressure can be adjusted based on the patient´s needs. The pressure can be separately adjusted in the Area A (seat section) or in the Area B (torso and legg mattress sections). The grey arrow below the Area A or Area B Pressure Level Icons indicates the optimized pressure.
Page 73
ALERTS UNPLUGGED When the power cable is unplugged or mains power fails, the screen will show the following alert and „POWER“ text is scrolling on the display. This alert will automatically disappear when mains power is restored. NOTE Red triangle with exclamation mark is displayed during this alert. Fig.
Page 74
COMPRESSOR (SCU) NOT CONNECTED When SCU is removed from the bed or communication between the bed and the SCU is lost this alert appears with „PUMP DISCONNECTED“ text. NOTE There is red triangle with exclamation mark on the display during this alert. To remove this alert: ►...
► Replace plastic handle every 4 years. ► Use Lateral Tilt carefully when lifting pole is installed on Eleganza 5 bed. To install the lifting pole: ► Insert lifting pole in corresponding sleeve fitting on lifting pole adapter at head end.
Manipulate carefully with head siderails when Hercules is installed! Hercules Patient Repositioner is intended for Hercules ready Eleganza 5 bed. Installation of Hercules must be done by qualified service technician authorized by the manufacturer. For detailed informations about Hercules follow the user manual for this product.
14.3 Infusion Stand WARNING! Risk of injury due to use of incorrect accessories or because of incorrect use! Infusion Stands must only be used for their intended use. Always read the instructions for use! ► Only mount an infusion pump to the lower (wider) telescopic section of an infusion stand above the head/foot end board.
14.4 Oxygen Bottle Holder WARNING! Risk of injury with oxygen bottle holder due to incorrect use or due to careless driving! ► Ensure the oxygen bottle holder is correctly fitted in correct position. ► It is necessary to place oxygen bottle holder (with or without O2 bottle) before transport to secure transport position.
Fig. Oxygen Bottle Holder in the adapter (position 4) 14.5 Ventilation Circuit Holder The ventilation circuit holder prevents an extubation. ► Always use Linet ® ventilation circuit holder to prevent extubation during any procedures. Applying ventilation circuit holder: ► Put ventilation circuit holder in hole on right or left side of the Backrest frame.
14.6 Writing Shelf The Writing Shelf is intended for writing of nursing staff. It is placed in the handles of the foot end (as on the picture). Fig. Writing Shelf 14.7 Monitor Shelf The Monitor Shelf is suitable for transporting monitors with a weight of up to 15 kg (33.07 lbs). Installing the monitor tray: ►...
Make sure the fall risk assessment was done properly before Protector use. The Protector is an optional accessory for the Eleganza 5 bed. The main purpose of the Protector is to reduce the risk of fall espe- cially at very risky patients (confused restless patients).
14.9 USB Connector WARNING! Risk of injury due to incorrect use! ► Ensure accessory pluged in USB connector is in pristine condition! User of the bed is responsible for the fact that this requirement is met. CAUTION! Risk of material damage due to incorrect use! ►...
■ by the EPA (Environmental Protection Agency) of the country in which the mattress replacement system is to be used. 15.1 Cleaning (Eleganza 5) Prepare for cleaning as follows: ►...
15.1.1 Daily Cleaning Clean the following bed parts: ■ All control elements for adjusting the bed ■ All handles □ CPR release handle ■ Bed ends ■ Siderails (in highest position) ■ Freely accessible mattress surface ■ Mobi-Lift® ■ Accessory rails 15.1.2 Cleaning before Changing Patients Clean the following bed parts: ■...
15.2 Cleaning (OptiCare) General guidance – Standard Cover: ► Do not use any strong acids or alkalines, (optimum pH range 6 – 8. Do not exceed pH of 9). Some hard surface cleaners have pH values outside this range, these are not suitable for use on coated textiles. ►...
Cleaning the SCU: ► Wipe SCU with disinfectant. ► Let SCU dry or wipe dry. 15.2.2 Full Cleaning and Disinfection Cleaning the mattress: ►, Deflate mattress and remove cover (see Removing the Mattress Cover). ► Check mattress cover top and base for any signs of damage. Replace or repair and completely disinfect mattress cover top and base if damaged.
► Do not use any strong acids or alkalines, (optimum pH range 6 – 8. Do not exceed pH of 9). ► Use only hospital-approved cleaners suitable for use on coated textiles and observe local directives concerning infection control. ► Do not use abrasive powders, steel wool, or other material and cleaning agents that might damage the mattress.
16 Troubleshooting DANGER! Risk of mortal injury due to electric shock! ► If a fault occurs, have the electric motor, power box or other electrical parts repaired by qualified personnel exclusively. ► Do not open the protective covers of the electric motor or the power box. Error/Fault Cause Solution...
Check all movable parts for wear. 17.2 Maintenance every 12 months Please refer to the Linet Periodic Preventative Maintenance and Safety Check Manual. 17.2.1 Spare Parts The product label is located on the inside of the longitudinal rail of the mattress platform frame. The product label contains informa- tion for claims and ordering replacement parts.
17.2.5 Electric Control Plug connections ► Replace O-rings on connectors. ► Check plugs connections for dirt and defects. Clean or replace if necessary. ► Check that the plug connectors are properly seated. Motors ► Check motor movement (adjust bed positions). Check for incorrect and interrupted movements.
® Our responsible Linet ® Service partners will ensure your Linet ® products are up and running when you need them. For more information on available service support and contract offerings, please contact us at podpora.servis@linet.cz and ask for techni- cal support.
Send the appliance (SCU) to the recycling of electrical appliances. 19 Warranty Linet ® will only be held responsible for the safety and reliability of products that are regularly serviced, maintained and used in accordance with the safety guidelines. Should a serious defect arise that cannot be repaired during maintenance: ►...
Page 93
20 Standards and Regulations 20.1 Eleganza 5 The Eleganza 5 bed comlies with the following standards: ■ IEC 60601-1 ■ CAN/CSA C22.2 NO. 60601-1 ■ ANSI/AAMI ES60601-1 ■ IEC 60601-1-2 ■ IEC 60601-1-6 ■ IEC 60601-2-52 ■ ISO 14971 20.2 OptiCare The OptiCare mattress replacement system complies with the following standards: ■ ...