Fig. Serial Label (WS17)
1.7 Acoustic signalisation
CONTINUOUS SOUND
BEEP + CONTINUOUS SOUND
REPEATED BEEP: 0,6s sound / 2,6s silence
MELODY: 3 beeps, pause, 2 beeps, longer pause, 3 beeps,
pause, 2 beeps
BEEP lasting 0,3s
4 TIMES REPEATED BEEP lasting 0,3s
BEEP lasting 0,5s
BEEP lasting 3s
BEEP lasting 5s
REPEATED BEEP during 3 minutes: 1,1s sound / 1,1s silen-
ce
10
SOUND
Scales Abbreviations
Max
maximum capacity of the weighing
instrument
Min
minimum capacity of the weighing
instrument
e
verification scale interval
T
tare value
MEANING
overheating
accumulator overcurrent
scales overload (only version with scales)
actuator overload
Siderail Signal (lateral tilt + head siderail or foot siderail down)
STOP error (all STOP buttons are disabled)
Bed Exit Alarm (only version with scales)
confirmation
stopping or locked function
lateral tilt 15° achieved
transition from tilt (lateral tilt, Trendelenburg, Antitrendelenburg)
to horizontal position
disconnected from the mains
positioning powered by the accumulator
start of service mode or end of service mode
keyboard error (positioning blocked)
system error
SCU disconnected (only if integrated mattress is used)
scale module disconnected (only version with scales)
Brake Signal (only version with Brake Signal)