Removing The Bed From Use; Deactivating The Accumulator - LINET Multicare LE Instructions For Use And Technical Description

Positionable bed for intensive care with scales and without scales
Hide thumbs Also See for Multicare LE:
Table of Contents

Advertisement

To cancel this status:
Press STOP button.
Status "Discharged accumulator"
The accumulator is regarded as discharged if the following condition is met:
Defined decrease of voltage depending on discharging current
This status is indicated by the accumulator status indicator flashing quickly.
The electric CPR position is the only possible position.
This status will be cancelled automatically when the bed switches to sleep mode.
To cancel this status:
Press STOP button.

10.3 Removing the Bed from Use

Remove the bed from service as follows:
Disconnect the bed from the mains.
Disconnect the ground wire.
Deactivate the accumulator.
Remove accessories.
To prevent damage during storage:
Pack or cover the bed and accessories.
Ensure that storage conditions are the same as the operating conditions.

10.4 Deactivating the Accumulator

To avoid damaging the bed and the environment during storage:
Deactivate the accumulator on the Attendant Control Panel.
To deactivate the accumulator on the supervisor:
Disconnect the bed from the mains.
Disconnect the ground wire.
Activate the keypad by pressing the GO button on the supervisor.
Press the Thighrest Up + Thighrest Down + Trendelenburg Position buttons at the same time and hold them for three
seconds.
The accumulator is deactivated.
To activate the accumulator again:
Connect Power Cable to the mains.
D9U001MC5-0101_08
33