The LED indicates the Accumulator charge status:
Yellow LED
Not lit
Short flashing (shortly lit, longer not lit) (circa 1.8 sec.)
Short flashing (0,2s lit, 0,2s not lit)
Long flashing (longer lit, shortly not lit) (circa 0.2 sec.)
Lit continuously for several hours (circa 10 hours), when bed is
connected to the mains.

10.2 Replacing the accumulator

CAUTION!
Damage to the bed due to incorrect accumulator replacement!
Have the accumulator replaced exclusively by qualified personnel.
Exclusively use accumulator approved by the manufacturer.
CAUTION!
Material damage due to overheating!
If the accumulator is faulty, degassing may occur. In rare cases this might cause deformations of the accumulator case,
control panel housing or cable.
Stop using the bed immediately (see Removing the Bed from Use).
Inform the manufacturer's service department.
CAUTION!
Risk of reducing accumulator durability due to incorrect use!
Use bed on accumulator only in crisis situations (e.g.: power blackout, patient complications during transport,
etc.)
After reconnecting bed to the mains charge accumulator to full capacity (see chart Accumulator charge status).
Have accumulator replaced exclusively by a qualified service organisation.
For more detailed information on how to replace the accumulator, request service manual from manufacturer.
Status "Faulty accumulator"
The accumulator is regarded as faulty if at least one of the following conditions applies:
Accumulator charging constantly
Low voltage on accumulator
Low charging current of accumulator
This status is indicated by the accumulator status indicator being constantly lit.
These statuses are summarised to Linis and written to Blackbox.
32
Accumulator charge status
Accumulator capacity is sufficient (charging completed)
Accumulator is charging - continue charging until the LED is
extinguished. In emer- gency cases, the accumulator can be
used as a backup power source for a short period. If LED is still
flashing after 12 hours of charging or stops flashing, but you
can not position with bed, accumulator is defective or broken.
Contact manufacturer.
Only CPR function can be used.
Low accumulator voltage - accumulator can not be used as a
backup power supply even for a short period; accumulator is
completely discharged or defective (if this type of signalisation
persists, it is necessary to replace the accumulator - service
action)
Accumulator absence or failure condition (accumulator is
connected incorrectly, line between the power supply and
accumulator is broken or accumulator fuses are faulty); contact
service department of the manufacturer in case of such signa-
lisation.