25.2. Reference unit mass estimated by weighing

Procedure:
• Enter <
• If the standard is to be weighed in a container, the container needs
to be put on the pan and tarred,
• Press
<Put standard>,
• Put the load on the pan. After stabilization the result is taken as
a standard (symbol
• At the same time the weight unit is changed to (w %).

25.3. Rederence unit mass inscribing into the memory

Procedure:
• Enter <
• Press
<Give piece mass> together with the screen keyboard,
• Enter a value and confirm by pressing
• At the same time the weight unit is changed to (w %).
26. WORKING MODES – COMPARATOR
The <
Comparator> working mode allows users to eatimate standard
deviation for a series of measurements. The standard deviation is
calculated according the sequence of measurements ABBA or ABA:
A – control mass standard,
B – tested mass standard.
Number of measurements for one series and the method: ABBA or ABA,
are set by users in parameter group <
Comparator> (see ch. 22).
<
Deviations> according to ch. 25.1 of this manual,
(Estimate standard mass), then a message is displayed:
). Confirm it by pressing
Deviations> according to ch. 25.1 of this manual,
(Give piece mass), then an editing field is displayed
,
Working modes> in submenu
77
,