21.1. Calibration procedure

• Enter submenu <
and select: "
• After entering the parameter the following message box appears:
• Take the load off the pan of platform 1,
• Press button
adjusting start mass: „Evaluation of starting mass",
• After the procedure has been completed the following message
box appears:
• Put calibration mass on pan of platform 1 and then select
Customer Calibration".
Customer Calibration> according to ch. 21
Platform 1 /
. The following message appears during
Calibration",