Median Filter; Ilter; Autozero - RADWAG WPY series User Manual

Hide thumbs Also See for WPY series:
Table of Contents

Advertisement

14.1. Median filter

The median filter is intended for eliminating short-lasting mechanical shocks.
Procedure:
• Enter <
Median Filter> and then set an appropriate value.
<
Accessible settings:
None - median filter is off
0.5, 1, 1.5, 2, 2.5 - filtering level to choose
14.2. Filter
This filter is intended to suppress continuous mechanical vibrations at the
cost of stabilization time.
Procedure:
• Enter <
Filter> and then set an appropriate value.
<
Accessible settings:
None, V.Fast, Fast, Average, Slow.
Notice:
The higher filtering level the longer stabilization time.

14.3. Autozero

The autozero function has been implemented in order to assure precise
indications. This function controls and corrects „0" indication.
While the function is active it compares the results continuously with
constant frequency. If two sequentional results differ less than the declared
value of autozero range, so the scale will be automatically zeroed and the
pictograms
Weighing> according to ch. 14 of the manual, select
Weighing> according to ch. 14 of the manual, select
and
will be displayed.
33