Switch Off Continuous Transmission In Current Unit; Set Lower Threshold; Set Upper Threshold; Read Lower Threshold - RADWAG WPY series User Manual

Hide thumbs Also See for WPY series:
Table of Contents

Advertisement

28.4.13. Switch off continuous transmission in current unit

Syntax: CU0 CR LF
Possible answers:
CU0_I CR LF
CU0_A CR LF

28.4.14. Set lower threshold

Syntax: DH_XXXXX CR LF, where: XXXXX – mass format
Possible answers:
DH_OK CR LF
ES CR LF

28.4.15. Set upper threshold

Syntax:
UH_XXXXX CR LF, where: XXXXX – mass format
Possible answers:
UH_OK CR LF
ES CR LF

28.4.16. Read lower threshold

Syntax: ODH CR LF
Possible answers: DH_MASA CR LF - command executed
Frame format:
1
2
D
H
- command comprehended but cannot be executed
- command comprehended and executed
- command executed
- command not comprehended (wrong mass format)
- command executed
- command not comprehended (wrong mass format)
3
4-12
space
mass
13
14
15
space
unit
100
16
17
18
space
CR
19
LF