Fault
After activating the power switch, the display and the view boxes do not light up.
Possible cause
Power supply to the device in-
terrupted.
The device fuse is defective.
Fault
Door does not open fully.
Possible cause
The device is switched off.
The door is under pressure.
Door mechanism defective.
Fault
The tray is not released when the door is opened.
Possible cause
The door is not fully open.
Fault
The tray cannot be locked.
Possible cause
There is a foreign object jam-
ming the locking mechanism.
The tray is jammed due to for-
eign objects under the table.
The tray is bent.
Fault
The interior lighting does not work.
Possible cause
The lighting element of the in-
terior lighting is defective.
17 March 2021
General faults
Measures
Check if the plugs are plugged in correctly.
Check the mains connection.
Change the fuse (see chapter 7.3.1 "Replacing
Fuses", page 90).
Measures
Switch on the device.
Close the door completely and then open it again.
Contact your local INFORS HT representative.
Measures
Push the door down to its bottom position.
Measures
Remove the foreign object.
Detach the table from its axis, tip it up and remove
the foreign bodies.
Replace the tray.
Measures
Replace the lighting element.
Multitron Standard - Operating Manual
Rectifying Faults
Personnel
Operator
Operator
Personnel
Operator
Operator
INFORS HT service tech-
nician or licensed dealer
Personnel
Operator
Personnel
Operator
Operator
Operator
Personnel
Qualified technicians