Procedure

6.9 Behaviour in Case of Interrupted Power Supply

17 March 2021
INFORMATION
The parameter setpoints are stored for approx. one month.
To switch off the device, proceed as follows:
1.
Deactivate all active parameters.
2.
Press the power switch on the top left of the casing.
3.
If the device is not going to be used for a longer period of time,
unplug it from the mains.
If the power supply to the device is interrupted during a cultivation
process (e.g. by flicking the power switch or in case of a power fail-
ure), all the parameters and timer setpoints will be stored, together
with the time remaining on the last active timer phase.
If the power supply is restored, the device will automatically restart
with the last stored setpoints. If a timer was active prior to the inter-
ruption to the power supply, the device will restart with the time re-
maining on the last active phase and the setpoints stored for this
phase.
P.FAIL
The alarm message
The alarm message can be confirmed by pressing any button; it will
then disappear.
Multitron Standard - Operating Manual
will appear in the view boxes as a warning
Operation
.
Page 81 of 114