Alarm message
Possible cause
The parameter has recently
been altered, and the required
setpoint has not yet been
reached.
Door not fully closed.
Capillary thermostat set too
low.
The ambient temperature does
not meet the requirements.
Cooling fins dusty.
Error during temperature meas-
urement
Heater fan not
working.
Alarm message
Possible cause
The parameter has recently
been altered, and the required
setpoint has not yet been
reached.
The ambient temperature does
not meet the requirements.
Error during temperature meas-
urement
Heater fan not working.
17 March 2021
The following alarm messages may occur:
Meaning
Temperature too low (actual value < setpoint - 1°C).
Measures
No action required. The alarm message will disap-
pear as soon as the setpoint is reached.
Close the door completely.
Set the capillary thermostat higher.
Check the ambient temperature on the device and
adjust if necessary.
Clean the cooling fins.
Check whether the Pt100 sensor is working
properly. In case of a defect, contact your INFORS
HT representative.
Contact your local INFORS HT representative.
Meaning
Temperature too high (actual value > setpoint + 1°C).
Measures
No action required. The alarm message will disap-
pear as soon as the setpoint is reached.
Check the ambient temperature on the device and
adjust if necessary.
Check whether the Pt100 sensor is working
properly. In case of a defect, contact your INFORS
HT representative.
Contact your local INFORS HT representative.
Multitron Standard - Operating Manual
Rectifying Faults
Personnel
Operator
Operator
Operator
Operator
Operator
INFORS HT service tech-
nician or licensed dealer
Personnel
Operator
Operator
INFORS HT service tech-
nician or licensed dealer
Page 83 of 114