Multitron Standard - Operating Manual
Operation
6.5.3

Programming a Cyclical Change

Page 74 of 114
INFORMATION
Do not set the timer to
clical operation.
The device will now run using the setpoints set for the first phase un-
til the set time elapses. The
.
the duration
As soon as the set time has expired, the system will change to the
Timer active
second phase. The
vice will now run with the setpoints set for the second phase until it
is switched off manually or the setpoints are modified.
INFORMATION
The timer function can be deactivated at any time by switching off
the timer. The device will then continue running with the last ac-
tive parameter settings.
Programming the timer for a cyclical change is a multi-step proce-
dure:
a)
Set the setpoints (RPM and/or °C) for the first phase using the
F button and activate the parameters.
b)
Set and activate the timer for the first phase using the F but-
ton.
c)
Set the setpoints (RPM and/or °C) for the second phase using
the FT button and activate the parameters.
d)
Set and activate the timer for the second phase using the FT
button.
ON
in the "FT" phase. This will activate cy-
Timer active
function symbol will flash for
function symbol will go out. The de-