Procedure
17 March 2021
Setting the Setpoints for Phase 1 (F Button)
1.
Press the F button (1) to select the parameter in question (RPM
or °C).
2.
Press the Plus or Minus button (2) to set the setpoint you want
to use for the selected parameter.
3.
Press the On/Off button (3) to activate the parameter.
4.
Repeat steps 1 – 3 as necessary for the other parameters.
INFORMATION
For details of how to set the parameters, see chapter 6.4 "Adjust-
ing, Activating and Deactivating Parameters", page 65.
Setting the Timer for Phase 1 (F Button)
1
5.
Select the timer function using the F button (3).
Timer
The
parameter symbol (2) will light up
6.
Use the Plus or Minus button (4) to set the required time.
7.
Activate the timer using the On/Off button (5).
Timer active
The
function symbol (1) will flash.
Multitron Standard - Operating Manual
Operation
1
2
2
4
3
.
3
5
Page 75 of 114