Permeation Oven Replacement - Thermo Scientific 43iQ Instruction Manual

Advertisement

Permeation Oven
Replacement
43iQ Instruction Manual
Use the following procedure to replace the permeation oven.
Equipment required:
Phillips drive, #2
Tube release tool (optional)
CAUTION Allow oven to cool down prior to servicing.
1. Turn the instrument OFF, unplug the power cord, and remove the
cover
(Figure
2–1).
2. Unplug DMC cables as shown.
3. Disconnect plumbing.
Figure 7–50. Perm Oven pt 1
L