Advanced Measurement
Thermo Scientific
The Advanced Measurement Settings screen allows the user to calibrate the
Settings
optical bench and set other advanced settings.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings
The Advanced Measurements Settings screen contains the following
information:
Optical Bench Settings: Calibrates detector gain, PMT supply, flash
G
voltage, and initial flash reference.
Flash Lamp: Enables/Disables flash lamp.
G
Optical Span Test: Enables/Disables test LED and reads the LED SO
G
concentration.
Extended Ranges: Enables/Disables extended ranges feature. When
G
changing to extended range from standard range, the PMP voltage
must be readjusted. For more information on readjusting the PMT
voltage, go to Settings>Measurement Settings>Advanced Measurement
Settings>Optical Bench Settings>Continue to PMT Settings.
Compensation: Allows the user to compensate for changes in
G
temperature and pressure concentration.
Perm Oven Selection: Allows the user to select the operating temperature
G
of the perm oven.
Pressure Calibration: Calibrates pressure.
G
Operation
Settings
2
3-71
43iQ Instruction Manual