Analog I/O
The Analog I/O screen provides status and alarms related to the analog
input/output module. If an item being monitored goes outside the lower or
higher alarm limit, an alarm is activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Analog I/O
The Analog I/O screen contains the following information:
Across:
G
G
G
G
G
G
Down:
G
G
G
G
G
G
Analog IO: This column lists items associated with the analog I/O.
Value: Displays the current value for each item.
Low Alarm: Displays low alarm status for each item.
High Alarm: Displays high alarm status for each item.
Units: Displays units for each item.
Voltage Output Channel 1–6: Displays real-time voltage output for
each channel.
Current Output Channel 1–6: Displays real-time current output for
each channel.
Chip Temperatures: Displays OK/Fail for chip temperatures.
Chip 1–3 Communication: Displays OK/Fail for each chip
communication.
Test Mode: Displays test mode on or off.
Operation
Settings
3-47
43iQ Instruction Manual