Operation
Settings
Concentration
3-36
43iQ Instruction Manual
The Concentration screen provides status and alarms for SO
concentration, background cal/checks, and span cal/checks. If the item
being monitored goes outside the lower or higher alarm limit, an alarm is
activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and
Alarms>Concentration
The Concentration screen contains the following information:
Across:
G
Concentration: This column lists items associated with the SO
G
concentration.
Value: Displays the current value for each item.
G
Low Alarm: Displays low alarm for each item.
G
High Alarm: Displays high alarm for each item.
G
Span Conc: Span concentration used in the span calibration or span
G
check.
Units: Displays units for each item.
G
Down:
G
SO
: SO
concentration.
G
2
2
Bkg Check Offset: Displays concentration based on the last
G
attempted background calibration. High alarm shows user defined
limit for acceptable background check offset.
2
2