The Calibrate Span Coefficient screen contains the following information:
Edit Span Concentration: User enters the span gas concentration when
G
in single range mode.
Edit High Range Span Concentration: User enters the high range span
G
concentration when in dual or auto range mode.
Edit Low Range Span Concentration: User enters the low range span
G
concentration when in dual or auto range mode.
Current High Range Concentration: Read only. Current high range
G
concentration reading when in dual or auto range mode.
Current Low Range Concentration: Read only. Current low
G
concentration reading when in dual or auto range mode.
Current Span Coefficient: Read only. Displays the current instrument
G
span coefficient.
Current High Range Span Coefficient: Read only. Displays the current
G
instrument high range span coefficient.
Current Low Range Span Coefficient: Read only. Displays the current
G
instrument low range span coefficient.
Calculated Span Coefficient: Read only. After the "Edit Span
G
Concentration" value is entered, the new calculated span coefficient is
displayed.
Calculated High Range Span Coefficient: Read only. After the "Edit
G
High Range Span Concentration" value is entered, the new calculated
high range span coefficient is displayed.
Calculated Low Range Span Coefficient: Read only. After the "Edit Low
G
Range Span Concentration" value is entered, the new calculated low
range span coefficient is displayed.
Calibrate: When pressed, updates the coefficient and the concentration
G
should match the span concentration.
Operation
Calibration
3-13
43iQ Instruction Manual