Valve Driver Outputs
Thermo Scientific
Table 9–7
lists the valve driver outputs.
Actual Output Voltage: 22−24 VDC
G
Maximum Current: 300 mA
G
Both positive and negative outputs are protected from over voltage and
G
over current by 500 mA fuses.
Figure 9–8. Valve Driver Outputs
Optional Equipment