Capillary; Removing The Capillary; Installing The Capillary - Ametek 3050-AMS User Manual

Moisture analyzer
Hide thumbs Also See for 3050-AMS:
Table of Contents

Advertisement

Figure 7-4.
Capillary location - 3050-
AMS.
7-4 | 3050 AMS Moisture Analyzer

Capillary

The capillary (Figure 7-4) is a part of the flow meter in the analyzer and
can easily be replaced when necessary.

Removing the capillary

1.
Remove the top cover from the analyzer and remove the cover from
the oven assembly. Facing the front of the Model 3050-AMS analyzer,
the capillary is found on top of the stream switching manifold (Figure
7-4).
3.
Disconnect the capillary and remove it by loosening both nuts and
gently pulling the tubing out of the stream switching manifold block.

Installing the capillary

1.
To install the capillary, gently push the tubing ferrules on the capillary
into the tubing on the manifold block until flush and tighten the nuts
to secure.
2.
Replace the oven lid and secure it with the two nuts. Replace the ana-
lyzer cover and secure it with the mounting screws.
Capillary
Proportinal
Solenoid
Valve
Manifold
Proportinal
Solenoid
Valve
Sensor