Dryer Compartment
Cover
Figure 7-6a.
Model 3050-AMS dryers, location.

Dryers

Removing the dryers

1.
At the rear of the analyzer, locate the dryer compartment on the right-
hand side (Figure 7-6a).
2.
Use a #2 Philips screwdriver to loosen the four captured screws on
the compartment cover.
3.
Remove the cover from the analyzer.
4.
The sample system (reference) dryer is located in the top position, and
the zero dryer is located in the bottom position.
5.
Taking one dryer at a time, use a 7/64" hex wrench to remove the
mounting screws that pass through the top of the dryer body and set
them aside.
6.
Loosen the 1/8 inch VCR fittings on the dryer and remove the dryer
from the analyzer. Discard the used VCR gaskets from the fittings.
Sample System Dryer
(top dryer)
Zero Dryer
(bottom dryer)
Dryer Mounting Screws
(2 each)
1/8" VCR Gasket
(2 ea dryer)
Service & Parts
|
7-7