7.7.2 DC ARC Calibration
DC ARC 5mA (3mA) Calibration Screen:
CALIBRATION
DC ARC 5mA(3mA)
CHANNEL: 1
OUTPUT
1.500kV
STANDARD
3.0mA
LOAD=500kΩ
1.
Press START to output voltage.
2.
When the status indicator shows PASS, it means the device has grabbed the ARC value
correctly and stopped the high voltage output.
7.8 When Calibration is Done
1.
Press SYSTEM to exit calibration mode or press▼ to return to A/D REFERENCE
calibration screen.
2.
To calibrate other channels, press Function Key [NEXT CH][PREV CH] to select another
channel for calibration.
READING
1.9mA
REMOTE LOCK
Calibration Procedure
NEXT CH
PREV CH
CORR
ERROR
7-23