This procedure should be performed in a temperature-stable environment after the pycnometer has
been allowed to warm for at least two hours. Before performing this procedure, check the chamber
cap to ensure that it is not the source of leaks. It should be free from particles, the O-ring should be
properly seated, and should not contain excessive grease.
C
L
HECK FOR
EAKS USING THE
Before performing this procedure, check the chamber cap to ensure that it is not the
source of leaks. It should be free from particles, the O-ring should be properly seated,
and it should not contain excessive grease.
Unit [n] > Diagnostics
Diagnostics using the Software on page 9 - 1
n
C
L
HECK FOR
EAKS USING THE
Before performing this procedure, check the chamber cap to ensure that it is not the
source of leaks. It should be free from particles, the O-ring should be properly seated,
and it should not contain excessive grease.
Check the cell and expansion chambers for leaks.
1. Allow the pycnometer to equilibrate thermally in a room with a stable temperature.
2. Press Alt + 1 to enter Manual mode.
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) — Jan 2021
Check the Cell and Expansion Chambers for Leaks
E
C
XPANSION
HAMBERS FOR
S
OFTWARE
K
EYPAD
A. Fill valve
B. Expansion valve
C. Vent valve
L
EAKS
12 - 7