Display Text
Volume of Cal Std:
(volume)
Which Chamber?
(chamber type)
Start:
(time) (date)
End:
(time) (date)
Cell[n] = (volume)
Dv[n] = (deviations)
or
Exp[n] = (volume)
Dv[n] = (deviations)
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) — Jan 2021
Description
Displays the volume of the calibration standard that was used for the
calibration. Press ENTER.
Select to view calibration data for the sample cell chamber or the
expansion chamber. Press CHOICE until the chamber is displayed,
then press ENTER.
The chamber type options are Cell volume and Expansion volume.
Displays the time and date the calibration began. Press ENTER.
Time: HH:MM:SS
Date: DD/MM/YY
Displays the time and date the calibration completed. Press ENTER.
Time: HH:MM:SS
Date: DD/MM/YY
Displays the Volume, depending on the chamber type selected (Cell
volume or Expansion volume).
[n] represents the cycle number.
Press CHOICE to exclude the displayed density (or volume) from the
calculated average. An asterisk next to the density indicates it has
been excluded. Press CHOICE again to remove the asterisk and have
it included.
Each time CHOICE is pressed to exclude or include the value. A new
deviation is calculated and displayed.
Press ENTER to view the values for the next cycle.
Press SAVE to return to the Reload prompt. Press SAVE to
automatically recalculate collected data and add all data reduction
messages back into the queue.
Review Calibration
10 - 19