Micromeritics ACCUPYC II 1345 Operator's Manual page 143

Gas displacement pycnometer
Hide thumbs Also See for ACCUPYC II 1345:
Table of Contents

Advertisement

11. Click
Continue
sages display until the calibration is complete.
12. Click
Done
to accept the new values, then click OK.
13. Do not remove the standard from the sample chamber until the operation is verified. See
Verify Operation on page 10 -
Calibrate Volume Scale
Field or Button
Analysis Conditions
Cell Volume
Continue [button]
Expansion Volume
Insert
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) — Jan 2021
to complete the second phase of the calibration (volume scale). Status mes-
11.
Description
Specify conditions for the calibration.
Click
Replace
to replace the values with those from an existing
Analysis Conditions file.
Displays the current cell volume (without an insert), or the volume of
the selected insert.
Place the standard into the sample cup and click
plete the volume scale calibration.
Displays the current expansion volume, and the date and time of cal-
ibration.
3
1 cm
unit
3
10 cm
unit
3
3
1.0 cm
100 cm
3
350 cm
and 2000 cm
If None is selected, place an empty sample cup in the chamber.
None, 0.1 cm
None, 3.5 cm
unit
None, 35 cm
3
units
None, Large, Small
Calibrate Volume Scale
Continue
to com-
3
3
3
3
, 10 cm
10 - 9