Hardware Information
Setting the 6-bit Configuration Switch
Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
Loss of data
C AUTI ON
Forced cold start erases all methods and data stored in the non-volatile
memory. Exceptions are calibration settings, diagnosis and repair log books
which will not be erased.
Save your methods and data before executing a forced cold start.
If you use the following switch settings and power the instrument up again, it will
start as described above.
Table 31
Boot Resident / Forced Coldstart
Agilent InfinityLab LC Series Diode Array Detectors User Manual
Boot Main System / Revert to Default Data
The instrument will boot to main mode and changes to the module's default
parameter. May be also required to load resident firmware into the module.
Boot Resident System / Revert to Default Data
The instrument will boot to resident mode and changes to the module's
default parameter. May be also required to load main firmware into the
module.
SW1
SW2
SW3
SW4
1
0
0
0
1
1
0
0
1
0
0
0
1
1
0
0
Note: The setting '0' (down) is essential.
SW5
SW6
Init Mode
0
0
Boot Main System / Keep Data
0
0
Boot Resident System / Keep Data
0
1
Boot Main System / Revert to Default Data
0
1
Boot Resident System / Revert to Default Data
201