7
Error Information
Detector Error Messages
Illegal Temperature Value from Sensor at Air Inlet
Error ID: 1072
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100 centigrade.
As a result the temperature control is switched off.
Probable cause
1
2
Heater at fan assembly failed
Error ID: 1073
Every time the deuterium lamp or the tungsten lamp (DAD only) is switched on or
off a heater self-test is performed. If the test fails an error event is created. As a
result the temperature control is switched off.
Probable cause
1
2
Heater Power At Limit
Error ID: 1074
The available power of the heater reached either the upper or lower limit. This
event is sent only once per run. The parameter determines which limit has been
hit:
0 means upper power limit hit (excessive ambient temperature drop).
1 means lower power limit hit (excessive ambient temperature increase).
Probable cause
1
Agilent InfinityLab LC Series Diode Array Detectors User Manual
The temperature sensor is defect.
Detector is exposed to illegal ambient condi-
tions.
Defective connector or cable.
Defective heater.
Excessive ambient temperature change.
Suggested actions
•
Replace the cable to the main board.
•
Please contact your Agilent service represen-
tative.
Verify that the ambient conditions are within the
allowed range.
Suggested actions
Please contact your Agilent service representa-
tive.
Please contact your Agilent service representa-
tive.
Suggested actions
Wait until temperature control equilibrates.
120