13
LAN Configuration

TCP/IP parameter configuration

TCP/IP parameter configuration

To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
The TCP/IP parameters can be configured by the following methods:
The LAN interface differentiates between several initialization modes. The
initialization mode (short form 'init mode') defines how to determine the active
TCP/IP parameters after power-on. The parameters may be derived non-volatile
memory or initialized with known default values. The initialization mode is
selected by the configuration switch, see
Agilent InfinityLab LC Series Diode Array Detectors User Manual
IP address
Subnet Mask
Default Gateway
by automatically requesting the parameters from a network-based DHCP
Server (using the so-called Dynamic Host Configuration Protocol). This mode
requires a LAN-onboard Module or a G1369C LAN Interface card, see
(DHCP)"
on page 212
by manually setting the parameters using Telnet
by manually setting the parameters using the Local Controller
Table 32
on page 209.
"Setup
207