8
Test Functions and Calibration
Cell Test
1 Run the Cell-Test with Agilent Lab Advisor (for further information see
Test Failed (low ratio value)
Cell Test Evaluation
Probable cause
1
2
Agilent InfinityLab LC Series Diode Array Detectors User Manual
Online-Help of user interface).
Figure 37
Cell Test – Results
Figure 38
Cell Test – Signals (example shows low intensity for flow cell)
Absorbing solvent or air bubble in flow cell.
Dirty or contaminated flow cell.
Suggested actions
Ensure the flow cell is filled with water, and free
from air bubbles.
Clean the flow cell as described in
Max-Light Cartridge Cell"
on page 165.
"Clean the
133